Assay Type | Sandwich-ELISA |
Analyte | IL-4 |
Format | 96T |
Reactivity | Human |
Regulatory Status | RUO |
Sensitivity | <3.0 pg/mL |
Standard Curve Range | 7.81 pg/mL-1000 pg/mL |
Assay Time | 2 hr |
Suitable Sample Type | For the quantitative determination of human IL-4 in Plasma and Serum. |
Sample volume | 50 μL |
ID | Components | Size |
CEA003-C01 | Pre-coated Anti-IL-4 Antibody Microplate | 1 plate |
CEA003-C02 | IL-4 Calibrator | 20 μg×2 |
CEA003-C03 | Biotin-Anti-IL-4 Antibody Concentrated Solution | 100 μL |
CEA003-C04 | Biotin Antibody Dilution Buffer | 8 mL |
CEA003-C05 | IL-4 Streptavidin-HRP Concentrated Solution | 0.5 mL |
CEA003-C06 | Streptavidin-HRP Dilution Buffer | 15 mL |
CEA003-C07 | 20× Washing Buffer | 50 mL |
CEA003-C08 | Sample Dilution Buffer | 15 mL×2 |
CEA003-C09 | Substrate Solution | 12 mL |
CEA003-C10 | Stop Solution | 6 mL |
ClinMax product are manufactured in a dedicated GMP facility and are compliant with our ISO 13485 quality management system.
The verification scheme was designed in accordance with the 9012 "Guidelines for the Verification of Quantitative Analysis Methods of Biological Samples" in the Chinese Pharmacopoeia 2020 Edition and the ICH M10 Guideline. Sensitivity, specificity, precision, recovery, and linearity were verified for multi-batch products and multiple concentration samples. See manual for more information on validation in DS.
Results are obtained by Log-Log Linear regression equation are used to draw the standard curve and calculate the sample concentration. The verification results indicate that this kit can be used for the quantitative determination of natural and recombinant human Interleukin 4 concentrations.
It has been calibrated against a highly purified human IL-4 and is evaluated with standard from NIBSC/WHO. Reference Reagent Interleukin-4 (Human, rDNA derived) NIBSC code: 88/656.
It is for research use only.
The opened kit should be stored per components table. The shelf life is 30 days from the date of opening.
For each experiment, each ELISA plate needs to set the standard curve. The minimum detectable concentration of CEA-C003 is less than 3.0 pg/mL.
Three human serum samples with high concentrations of IL-4 were diluted 1:2, 1:4, 1:8, 1:16, 1:32, 1:64 with Dilution Buffer to produce samples with values within the dynamic range and then assayed. On average, 103.83% of IL-4 was detected from serum samples.
Ten replicates of each of five samples containing different IL-4 concentrations were tested in one assay , Intra-Assay Precision CV<10%.
Five samples containing different concentrations of IL-4 were tested in three independent assays , Inter-Assay Precision CV<15%.
Recombinant IL-4 was spiked into 5 human serum samples, and then analyzed. On average, 101.42% of IL-4 was recovered from serum samples.
Recombinant IL-4 was spiked into cell supernatant sample, and then analyzed. On average, 97.94% of IL-4 was recovered from serum samples.
Bilirubin (simulated jaundice) concentration should be less than 20mg/dL, hemoglobin (simulated hemolysis) concentration should not be higher than 3500mg/dL, triglyceride (simulated lipemia) concentration may not be higher than 2.0g/L, Heparin concentration should be less than 40U/mL, EDTA concentration should be less than 4mg/mL, and Sodium citrate concentration may not be higher than 40mg/mL, it does not affect the detection result.
Please contact us via TechSupport@acrobiosystems.com if you have any question on this product.
ComboX 는 ACROBiosystems에서 새롭게 출시한 범용 제품으로 Streptavidin 계열 제품, 코팅 플레이트, 항체 및 융합 단백질 약물 아이소타입 컨트롤, 범용 항체 등 일련의 범용 제품을 조합하여 약물 연구 개발에 도움이 되는 브랜드입니다!
최신 Lambda 돌연변이, Omicron 돌연변이 및 기타 중요한 돌연변이 제품이 출시되었습니다. 이 제품은 핫스팟 단백질 및 주요 고주파 돌연변이를 포함하며 돌연변이 사이트 디자인은 PANGOLIN을 참고하며 다양한 태그를 선택할 수 있습니다.
고품질 GMP 등급 사이토카인 IL-15,IL-7, IL-21 등 제품은 신속한 임상/신약승인신청을 위해 개발된 것입니다.
50+ car-t의 인기 타겟을 커버하고, car 발현 검사를 위해 특별히 설계된 항원 단백질은 스트림 검증을 통해 고특이성 car 발현을 검출할 수 있으며, 고 배치간 일관성을 가지고 있다.pe/fi표기 단백질로 한 단계 염색하면 고특이성, 무 백그라운드로 car 발현 검사를 검사할 수 있습니다.
GMP 등급 사이토카인. 고품질 세포 활성화 및 확장 시약. 유전자 편집 시약, 효소 CAR 검사 시약 등 제품. 세포 및 유전자 요법 고객님에게 포괄적인 솔루션을 제공하고 초기 약물 발견에서 임상 연구에 이르기까지 모든 단계에서 귀하와 동반하겠습니다.
cd20, claudin18.2, cd133, gprc5d, ccr5, ccr8을 포함한 안정적이고 활성이 높은 전장 다중 막관통 단백질은 면역, elisa, spr, bli, 세포 실험, car 양성률 검사 등에 광범위하게 사용된다. vlp, 스케일제거제, nanodisc의 다양한 기술 플랫폼은 다중 막관통 단백질을 목표로 하는 약물 연구에 도움을 준다.
cd3, pd-1, ctla-4 등 70+ 인기 이중 항체 타겟을 커버하며 활성이 높고 특이성이 강하며 배치간의 일관성이 좋아 항체면역과 선별과정에 적합하다.
현재 이미 알려진 거의 대부분의 면역 체크 포인트 분자를 포함하여 다양한 레벨과 물종을 제공하여 고객이 선택할수 있도록 하고있다. 천연중합체 형식은 mals를 통해 검증되었고 생물활성은 elisa/spr/bli/facs 등을 통해 검증되었으며 또한 biotin/fitc 표지는 높은 처리량을 가진 항체를 선별하는데 편리하다.
ACROBiosystems는 ADC약물의 연구개발을 돕기 위해 노력하고 있으며, 다양한 인기 타겟에 서로 다른 종, 다른 라벨의 제품을 개발했으며, 높은 순도, 높은 친화력 등 특징을 가지고 있으며, 면역, 항체 선택, spr, 세포 활성 검사 및 기타 실험에 적합하고, 해당 protocol을 무료로 제공합니다.
Fc 수용체 단백질은 모든 분자를 포함하고 있을 뿐만 아니라, 일반적인 돌연변이와 바이오틴 표기 종류을 포함하고 있어 고객들이 단일 클론 항체 개발을 가속화할 수 있다.
포괄적으로 풍부한 사이토카인 표적 단백질은 HEK293 진핵생물 시스템에 의해 발현되며 천연 구조에 더 가깝다. 고순도는 SDS-PAGE/HPLC/MALS에 의해 검증된다. 높은 생물학적 활성이 ELISA/SPR/BLI에 의해 검증된다. 배치간 일관성이 우수하여 항체 면역, 스크리닝 및 품질 관리 과정에 적합하다.
Aneuro는 ACROBiosystems가 뇌과학 연구에 초점을 맞춘 제품 라인으로서 치료 및 진단 연구 단백질, PFFs와 재조합 신경 인자 등 뇌과학 연구 분야에 중요한 단백질 제품을 제공하여 뇌과학 연구에 도움이 된다.
This web search service is supported by Google Inc.